作者单位:华中理工大学自动控制工程系 (瞿安连,李玲),华中理工大学自动控制工程系(康华光)
摘    要:本文首先简要讨伦磁共振成象(MRI)系统在心脏成象中产生伪影的原因及消除伪影的方法,然后着重论述MRI心脏成象门控仪的设计原则和工作原理.临床实验结果表明,心脏成象门控仪的应用使心脏MRI图象具有较高的诊断价值.

关 键 词:磁共振  成象  心脏门控  运动伪影  浮地放大器  光导纤维

On MR Cardiac Gating Imaging
Qu Anlian Li Lin Kang Huaguang.On MR Cardiac Gating Imaging[J].JOURNAL OF HUAZHONG UNIVERSITY OF SCIENCE AND TECHNOLOGY.NATURE SCIENCE,1990(3).
Authors:Qu Anlian Li Lin Kang Huaguang
Institution:Qu Anlian Li Lin Kang Huaguang
Abstract:Magnetic Resonance Imaging (MRI) is useful for cardiovascular system examination. In conventional MRI techniques several minutes are needed to sample and produce an image. Consequently, when imaging sequences are made, as the data are sampled at random from systolic and diastolic phases of the cardiac cycle, motion artifacts and edge unsharpress occurs in moving structures of the image. An electrocardiograph (ECG) sensing and gating device has been developed to sense the ECG signal and generate a sequence of triggering pulses to synchronize the imaging sequence with a specific phase of the cardiac cycle. Noises and artifacts induced by the radio-frequency pulse and gradient magnetic fields inherent in MRI are electrically suppressed with a floating-earth amplifier and a fibre-optic cable to amplify and transmit the ECG signal. This device is currently satisfactorily used for medical examination in Beijing Cancer Hospital.