Minimum inductance optimal design for the gradient coil
Authors:Chen Yin-ping  Zhu Guo-qiang
Institution:(1) School of Electronic Information, Wuhan University, 430072 Wuhan, Hubei, China
Abstract:In MRI (Magnetic Resonance Imaging), a crucial role of gradient coils is to image organism, meanwhile the inductance of the coils determines the speed of imaging. So it is of great importance to optimize designs of the gradient coils. The target field approach is an effective method to design the gradient coils. Having applied this method and performing many numerical tests, we achieved the designs of the x-、y-、z- gradient coils , with the linearity of the fields generated by the coils in a sphere of radius 0.30 m less than 5%, in which the inductance and resistance of the coils also meet the requirements.
Keywords:magnetic resonance imaging  linearity  target field  Fourier transform