摘    要:The effect of geometric parameters of the coil element in an array on signal-to-noise ratio (SNR) and g factor of parallel imaging are analyzed, and the optimal parameters for the maximal SNR and minimal g factor are also given in this paper. In addition, the effects of coupling on SNR and g factor have been analyzed. The test results indicate that g factor for parallel imaging is greatly increased and SNR is decreased for coupling. According to the optimization, a six-channel parallel coil has been designed for torso imaging.

关 键 词:MRI   phased array coil   SENSE   parallel imaging.

Simulation of g factor for size optimization and coupling in phased array coil design
SONG Xiaoyu,ZHANG Hongjie,WANG Le,BAO Shanglian,SUN Yanrong,ZU Donglin. Simulation of g factor for size optimization and coupling in phased array coil design[J]. Progress in Natural Science, 2006, 16(2): 120-124. DOI: 10.1080/10020070612331343202
Authors:SONG Xiaoyu  ZHANG Hongjie  WANG Le  BAO Shanglian  SUN Yanrong  ZU Donglin
Affiliation:Beijing Key Laboratory of Medical Physics & Engineering, Peking University, Beijing 100871, China;Physics School, Peking University, Beijing 100871, China
Abstract:The effect of geometric parameters of the coil element in an array on signal-to-noise ratio (SNR) and g factor of parallel imaging are analyzed, and the optimal parameters for the maximal SNR and minimal g factor are also given in this paper. In addition, the effects of coupling on SNR and g factor have been analyzed. The test results indicate that g factor for parallel imaging is greatly increased and SNR is decreased for coupling. According to the optimization, a six-channel parallel coil has been designed for torso imaging.
Keywords:MRI  phased array coil  SENSE  parallel imaging