Consistency Conditions for Cone-Beam CT Data Acquired with a Straight-Line Source Trajectory
Authors:LEVINE+Margo+S  SIDKY+Emil+Y  PAN+Xiaochuan+Department+of+Radiology  University+of+Chicago  +S.Maryl  +Ave.  Chicago  IL+  USA
Affiliation:LEVINE+Margo+S,SIDKY+Emil+Y,PAN+Xiaochuan+Department+of+Radiology,University+of+Chicago,5841+S.Maryl,+Ave.,Chicago,IL+60637,USA
Abstract:A consistency condition is developed for computed tomography(CT) projection data acquired from a straight-line X-ray source trajectory.The condition states that integrals of normalized projection data along detector lines parallel to the X-ray path must be equal.The projection data is required to be untruncated only along the detector lines parallel to the X-ray path,a less restrictive requirement compared to Fourier conditions that necessitate completely untruncated data.The condition is implemented numeri...
Keywords:data redundancy  John's equation  consistency conditions