首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Detection of human fibrin by immunological piezoelectric quartz crystal microbalance technique
Authors:Pei Renjun  Hu Jiming  Hu Yi  Zeng Yun’e
Institution:(1) Department of Chemistry, Wuhan University, 430072 Wuhan, China;(2) Department of Analysis-Measurement Science, Wuhan University, 430072 Wuhan, China
Abstract:The detection of human fibrin has been performed by immunological piezoelectric quartz crystal microbalance technique. Anti-fibrin antibody was immobilized onto the gold electrodes of a 9 MHz AT-cut piezoelectric quartz crystal with bovine serum albumin adsorption, followed by glutaraldehyde cross-linking. The antibody-bound crystals had good response to human fibrin in the concentration range of 10−4∼10−1g/L. The sensor system had good selectivity. The crystal could be regenerated nearly 10 times without detectable decrease of sensitivity. Supported by The National Natural Science Foundation of China Pei Renjun: born in Nov. 1971, Ph. D. graduate student
Keywords:human fibrin  quartz crystal microbalance  antibody immobilization  regeneration