首页 | 本学科首页   官方微博 | 高级检索  
     检索      

基于液滴分析技术检测水中的溶解氧及盐度
引用本文:冯国红,裘祖荣,廖和琴,张亚娟.基于液滴分析技术检测水中的溶解氧及盐度[J].天津大学学报(自然科学与工程技术版),2012(12):1083-1088.
作者姓名:冯国红  裘祖荣  廖和琴  张亚娟
作者单位:天津大学精密测试技术及仪器国家重点实验室;东北林业大学工程技术学院
基金项目:中央高校基本科研业务费专项资金资助项目(DL12BB27)
摘    要:为使液滴分析技术定性识别液体时能定量给出主要的水质指标,将光纤溶解氧传感器引入液滴分析技术,利用荧光淬灭原理检测溶解氧含量;同时,利用盐度与液滴指纹图的关系建立拟合方程,检测盐度.将溶解氧探头放入盛水的烧杯中与液滴结构中进行对比实验,两者最大相差0.37%,说明在液滴结构中检测溶解氧具有可行性;对不同盐度的溶液进行实验,结果表明随着盐度的增加光纤电压值增大,建立了盐度的拟合方程;通过检验发现式(2)效果较好,最大误差为5.5%.

关 键 词:液滴分析技术  溶解氧  盐度  液滴指纹图  荧光淬灭

Droplet Analysis Technology in Detection of Dissolved Oxygen and Salinity in Water
FENG Guo-hong,QIU Zu-rong,LIAO He-qin,ZHANG Ya-juan.Droplet Analysis Technology in Detection of Dissolved Oxygen and Salinity in Water[J].Journal of Tianjin University(Science and Technology),2012(12):1083-1088.
Authors:FENG Guo-hong  QIU Zu-rong  LIAO He-qin  ZHANG Ya-juan
Institution:1(1.State Key Laboratory of Precision Measuring Technology and Instruments,Tianjin University,Tianjin 300072,China;2.College of Engineering and Technology,Northeast Forestry University,Harbin 150040,China)
Abstract:To quantitatively measure water quality parameters when droplet analysis technology is applied to identify-ing the liquid, fiber dissolved oxygen sensor was introduced into droplet analysis technology. Fluorescence quench-ing principle was adopted to detect dissolved oxygen concentration. Meanwhile, based on the relationship between salinity and liquid droplet fingerprint (LDF), the fitting equation is established to inspect the salinity. The compara- tive experiments were conducted by putting the probe into the breaker and the dissolved oxygen droplet structure, respectively, and the results show that the maximum error is 0.37%, which means that it is feasible to detect the con- centration of dissolved oxygen in the droplet structure. Experiments were performed on liquids with different salinities, and the results show that the optical fiber voltage increases with the increase of salinity. The fitting equa-tion of salinity was established, and it is shown through inspection that the fitting equation is significant and the maximum error is 5.5%.
Keywords:droplet analysis technology  dissolved oxygen  salinity  liquid drop fingerprint (LDF)  fluorescencequenching
本文献已被 CNKI 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏