摘    要:建立了测定新药UP0044中甲醇、乙醇、二氯甲烷、乙酸乙酯、甲苯、四氢呋喃、乙酸7种有机溶剂含量的方法,以确定它们在药物中的残留量和比例.实验方法采用毛细管气相色谱法,氢火焰离子检测器(FID),以DB-624色谱柱(30m×250pm×1.4μm)对样品进行分离,采取程序升温,载气为氦气,外标法计算含量.结果显示:在各色谱条件下,7种溶剂的线性良好(R^2〉0.99),平均回收率为90%~110%,RSD为1.2%~8.8%,最低检出限0.9×10^-8~1.69×10^-4;所测试的4批样品中除甲醇、乙醇有检出外,其他溶剂均未检出.

关 键 词:UP0044  有机溶剂残留量  气相色谱法
文章编号:1671-1114(2008)01-0028-05
收稿时间:2007-10-18
修稿时间:2007-10-18

Determination of residual organic solvents in a new pharmaceutical-developing(UP0044) by capillary gas chromatography
SONG Wenhua,SUN Ting,GAO Minling,JIN Xinglong,WANG Guofeng. Determination of residual organic solvents in a new pharmaceutical-developing(UP0044) by capillary gas chromatography[J]. Journal of Tianjin Normal University(Natural Science Edition), 2008, 28(1): 28-32
Authors:SONG Wenhua  SUN Ting  GAO Minling  JIN Xinglong  WANG Guofeng
Affiliation:S0NG Wenhua, SUN Ting, GA0 Minling, J IN Xinglong, WANG Guo feng (1. School of Environmental Science and Safety Engineering, Tecnology of Tianjin University, Tianjin 300191, China; 2. Research Center of Environmental Science and Safety Engineering, Tianjin Polytechnic University, Tianjin 300160, China; 3. School of Environmental Science and Engineering, Shanghai Jiaotong University, Shanghai 200030, China; 4. Department of Life Science, SGS-CSTC Standards Technical Serxices Co. , Ltd Shanghai Branch, Shanghai 200233, China)
Abstract:A method for determining contents of seven kinds of residual organic solvents such as methanol, ethanol, methylene chloride, ethyl aetate, toluene, tetrahydrofuran and acetic acid in a new pharmaceutical-developing (UP0044) was established by capillary gas chromatography. The experimental conditions were DB-624 silica capillary column, FID detector, carrier helium gas and external standard calibration method. The results indicated that the seven residual solvents were separated completely and the linearity were good(R2〉0.99). Average recovery was 90%-110%, RSD was 1.2%-8.8%. The lowest detection limit was 0. 9×10^-6-1.69× 10^4. Both methanol and ethanol solvents were detected in 4 batches of the UP0044, but other solvents was no done.
Keywords:a new pharmaceutical-developing(UP0044)  residual organic solvent  gas chromatography