Reliable Remote-Monitoring Electrochemical Potentiostat for Glucose Measurements
Authors:JIN Yang  WANG Hong  LV Zhengliang  YANG Shiyuan  CAI Haoyuan  JIANG Junfeng
Institution:aDepartment of Automation, Tsinghua University, Beijing 100084, China;bSiemens Ltd. Corporate Technology, Beijing 100102, China
Abstract:Electrochemical methods have been widely used in the chemical and pharmaceutical industries, which require accurate concentration measurements, chemical reaction detections and analyses. The electrochemical potentiostat, the core element in electrochemical instruments, have been discussed as a hot topic addressing the difficulty of applying high-precision constant voltage and picoampere current measurements. Meanwhile, reliable potentiostats are in demand for complicated industrial environments with noises as well as requirements of remotemonitors. This paper describes a potentiostat for industrial glucose measurement that is not only accurate but also fault tolerant to guarantee high reliability in industrial environments. The instrument uses standard industrial communication protocols, profibus, and a 4–20 mA current loop, for remote control and monitoring. Experimental results show that this design has 0.01% accuracy with 1 mV resolution for voltage applications and 0.01% accuracy with 1 pA resolution for current measurements. The design is also shown to be highly reliable in noisy environments.
Keywords:electrochemical potentiostat  reliable  remote-monitoring  glucose measurements