首页 | 本学科首页   官方微博 | 高级检索  
     

A new simulated annealing method of designing NMR biplanar shim coils
引用本文:QI Feng,TANG Xin,WANG Weimin,SHEN Yifei,LIU Shaopo,JIANG Zhongde. A new simulated annealing method of designing NMR biplanar shim coils[J]. 自然科学进展(英文版), 2006, 16(7): 747-752. DOI: 10.1080/10020070612330063
作者姓名:QI Feng  TANG Xin  WANG Weimin  SHEN Yifei  LIU Shaopo  JIANG Zhongde
作者单位:1.Key Laboratory for Quantum Information and Measurements of Ministry of Education, School of Electronics Engineering & Computer Science, Peking University, Beijing 100871, China; 2. Institute of Heavy Ion Physics, and Beijing Key Laboratory of Medical Physics and Engineering, Physical School, Peking University, Beijing 100871, China
基金项目:国家高技术与发展计划(863计划);北京市科委资助项目
摘    要:A new SA (simulated annealing) method is presented for the design of longitudinal shim coils for NMR (nuclear magnetic resonance) equipment. First, AGL (adaptive Gauss-Lobatto) quadrature used by the traditional SA method is analyzed. Two kinds of complete elliptic integrals are introduced to integrate the spatial magnetic field, which increases the computing speed by 13.3 times. Then, the system energy calculation is extended to the field of the z-axis. The results illustrate that the coil-set efficiency designed by the new SA method is 15% higher than that of the traditional one, and the field inhomogeneity is reduced from 1.20% to 0.49%. Finally, a prototype of coil assembly for NMR rock-core analyzer is built, and the experimental results are in good agreement with the theoretically calculated results.

关 键 词:NMR   shim coil design   simulated-annealing method   complete elliptic integral.

A new simulated annealing method of designing NMR biplanar shim coils
QI Feng,TANG Xin,WANG Weimin,SHEN Yifei,LIU Shaopo,JIANG Zhongde. A new simulated annealing method of designing NMR biplanar shim coils[J]. Progress in Natural Science, 2006, 16(7): 747-752. DOI: 10.1080/10020070612330063
Authors:QI Feng  TANG Xin  WANG Weimin  SHEN Yifei  LIU Shaopo  JIANG Zhongde
Affiliation:1. Key Laboratory for Quantum Information and Measurements of Ministry of Education, School of Electronics Engineering & Computer Science, Peking University, Beijing 100871, China
2. Institute of Heavy Ion Physics, and Beijing Key Laboratory of Medical Physics and Engineering, Physical School, Peking University, Beijing 100871, China
Abstract:A new SA (simulated annealing) method is presented for the design of longitudinal shim coils for NMR (nuclear magnetic resonance) equipment. First, AGL (adaptive Gauss-Lobatto) quadrature used by the traditional SA method is analyzed. Two kinds of complete elliptic integrals are introduced to integrate the spatial magnetic field, which increases the computing speed by 13.3 times. Then, the system energy calculation is extended to the field of the z-axis. The results illustrate that the coil-set efficiency designed by the new SA method is 15% higher than that of the traditional one, and the field inhomogeneity is reduced from 1.20% to 0.49%. Finally, a prototype of coil assembly for NMR rock-core analyzer is built, and the experimental results are in good agreement with the theoretically calculated results.
Keywords:NMR  shim coil design  simulated-annealing method  complete elliptic integral
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《自然科学进展(英文版)》浏览原始摘要信息
点击此处可从《自然科学进展(英文版)》下载全文