非接触动态心电监测系统研究
引用本文:李文宝,陈贤祥,胡欣宇,任仁,夏善红.非接触动态心电监测系统研究[J].科学技术与工程,2013,13(21):6061-6065.
作者姓名:李文宝  陈贤祥  胡欣宇  任仁  夏善红
作者单位:中国科学院电子学研究所,中科院电子所,中科院电子所,中科院电子所,中科院电子所
基金项目:国家高技术研究发展计划(863计划)
摘    要:针对传统心电监测电极需要与皮肤直接接触易,引起皮肤不适且准备时间长的问题,设计了一种新型非接触动态心电监测系统。该系统基于电容耦合原理,采用三同轴电缆屏蔽驱动与接地屏蔽技术,有效地提高了系统的输入阻抗。建立了电极等效电路模型,分析了噪声的关键影响因素,对电路的结构与参数进行了优化分析。实验结果表明,该系统能够在静止和非剧烈运动两种状态下非接触测量心电波形中QRS波和T波,实现心电信号的非接触有效测量。

关 键 词:心电监测  非接触  电极  电容耦合  三同轴电缆
收稿时间:2013/3/26 0:00:00
修稿时间:2013/4/12 0:00:00

Research on Noncontact Dynamic ECG System
LI Wen-bao,CHEN Xian-xiang,HU Xin-yu,REN Ren,XIA Shan-hong.Research on Noncontact Dynamic ECG System[J].Science Technology and Engineering,2013,13(21):6061-6065.
Authors:LI Wen-bao  CHEN Xian-xiang  HU Xin-yu  REN Ren  XIA Shan-hong
Institution:1(State Key Lab of Transducer Technology,Institute of Electronics,Chinese Academy of Sciences 1,Beijing 100190,P.R.China;Graduate University of the Chinese Academy of Sciences 2,Beijing 100039,P.R.China)
Abstract:Electrode of traditional ECG monitoring system need direct contact with skin using conductive paste, which prepare for a long time, besides, the conductive paste may cause allergy of skin. To solve above problem, a novel kind of Noncontact ambulant ECG monitoring system based on capacitance coupling principle was designed in this paper. A triax cable shielding driven and ground shielding technology was adopted to increase the input impedance of system. ECG can be detected through clothing. The ECG signal noise was analyzed and signal detection circuit parameters were optimized based on the equivalent model of electrode. Experiment results shows that the QRS and T waves can be clearly seen both at quiet and walking state, which implement ECG noncontact measurement effectively.
Keywords:ECG  non-contact  electrode  capacitance coupling  triax cable