有机物种类对恒电位电解法测定COD值准确性的影响
引用本文:王西志,莫恒亮,方耀,唐阳,陈咏梅,杨晓进,万平玉.有机物种类对恒电位电解法测定COD值准确性的影响[J].北京化工大学学报(自然科学版),2015,42(4):10.
作者姓名:王西志  莫恒亮  方耀  唐阳  陈咏梅  杨晓进  万平玉
作者单位:北京化工大学理学院新危险化学品评估及事故鉴定国家级中心实验室,北京,100029;杭州电化集团有限公司,杭州,311228
基金项目:中央高校基本科研业务费专项
摘    要:以β-PbO2/石墨电极为工作电极,在支持电解溶液中加入一定量待测有机物,在低于析氧电位下进行恒电位电解时,因电极表面吸附态羟基自由基的消耗速率受电解质溶液中有机物分子向电极表面的扩散过程控制,表现为阳极电流强度的增加值(ΔI)与有机物质量浓度ρ之间呈线性关系。本文研究了多种待测有机物的ΔI与ρ的线性关系,发现由于有机物种类不同造成羟基自由基消耗速率不同,因而线性比例系数有所不同。实验结果表明在测定混合水样的化学需氧量(COD)值时,应采用与待测水样具有相似有机物混合比例的标准溶液制作工作曲线,以减少测量误差。

关 键 词:β-PbO2/石墨电极  化学需氧量(COD)  恒电位电解  阳极电流强度  羟基自由基
收稿时间:2014-12-29

The effect of chemical species on the accuracy of COD values measured by constant potential electrolysis method
WANG XiZhi,MO HengLiang,FANG Yao,TANG Yang,CHEN YongMei,YANG XiaoJin,WAN PingYu.The effect of chemical species on the accuracy of COD values measured by constant potential electrolysis method[J].Journal of Beijing University of Chemical Technology,2015,42(4):10.
Authors:WANG XiZhi  MO HengLiang  FANG Yao  TANG Yang  CHEN YongMei  YANG XiaoJin  WAN PingYu
Institution:1. National Fundamental Research Laboratory of New Hazardous Chemicals, School of Science, Beijing University of Chemical Technology, Beijing 100029; 2. Hangzhou Electrochemical Group Co. Ltd., Hangzhou 311228, China
Abstract:When the organic matter in aqueous solution is electrolyzed at constant potential using a β-PbO2/graphite electrode as the working anode, the reaction rate between organic molecules and adsorbed hydroxyl radicals is controlled by the diffusion of the organic molecules towards to the surface of the electrode. For fined stirring rate and viscosity, the variation in anodic current has a linear relationship with the concentration of organic matter in water. Different kinds of chemical species were tested in this study. Although each chemical species displayed a good linear relationship, the proportionality coefficients were different due to the different diffusion coefficients. Therefore, when a sample containing a mixture of chemical species is tested, a calibration curve should be made based on standard solutions with similar proportions of these chemicals.
Keywords:
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《北京化工大学学报(自然科学版)》浏览原始摘要信息
点击此处可从《北京化工大学学报(自然科学版)》下载免费的PDF全文