单帧单曝光图像法对粉雾剂粉末模型分散效果的研究
引用本文:邹珺,廖跃华,刘阳,周骛,陈岚.单帧单曝光图像法对粉雾剂粉末模型分散效果的研究[J].上海理工大学学报,2018,40(2):166-170,176.
作者姓名:邹珺  廖跃华  刘阳  周骛  陈岚
作者单位:上海理工大学医疗器械与食品学院;上海健康医学院医疗器械学院;上海理工大学能源与动力工程学院
基金项目:上海市科委产学研医合作项目(14DZ1930500)
摘    要:利用单帧单曝光图像法(SFSEI),从相互垂直的2个方向分别对不同流量的气流分散作用下的粉雾剂粉末模型的分散粒径(及粒径分布)和颗粒速度(及速度分布)进行测量;将结果与激光衍射法、数值模拟法进行比较;分析了SFSEI法在粉雾剂分散效果测量中的适用性。研究结果表明:实验中颗粒统计数量在750以上时,所得平均粒径和平均速度结果无明显差异;对于实验条件和流道结构,竖直方向拍摄比水平方向拍摄得到的结果更接近激光衍射法和数值模拟的结果;同时,颗粒速度的测量会受粒径大小的影响。使用SFSEI方法时应尽可能减少离焦颗粒比例以提高测量精度,在气体流量较低的情况下更易于实现较高的测量准确性。

关 键 词:粉雾剂  分散效果  单帧单曝光图像法  粉末
收稿时间:2017/3/7 0:00:00

Analysis on the Dispersion Effect of a Powder Model for Dry Powder Inhalers by the Single Frame and Single Exposure Image Method
ZOU Jun,LIAO Yuehu,LIU Yang,ZHOU Wu and CHEN Lan.Analysis on the Dispersion Effect of a Powder Model for Dry Powder Inhalers by the Single Frame and Single Exposure Image Method[J].Journal of University of Shanghai For Science and Technology,2018,40(2):166-170,176.
Authors:ZOU Jun  LIAO Yuehu  LIU Yang  ZHOU Wu and CHEN Lan
Institution:School of Medical Instrument and Food Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China,School of Medical Instrument, Shanghai University of Medicine and Health Sciences, Shanghai 201318, China,School of Medical Instrument and Food Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China,School of Energy and Power Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China and School of Medical Instrument and Food Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China
Abstract:At different gas flow velocities and from two perpendicular directions, the particle size distribution and particle velocity distribution of a powder model for dry powder inhalers were investigated using the single frame and single exposure image method (SFSEI). The results were compared with those by the laser diffraction and numerical simulation methods. The applicability of SFSEI in the measurement of dry powder dispersion was analyzed. It is indicated that the measuring results are not sensitive to the particle quantity when it is no less than 750; the data obtained from vertical images are closer to the ones from the laser diffraction and the simulation results; the particle velocity measurement may be affected by the particle size. Based on this, it is shown that when applying the SFSEI method, the quantity of particles which are out of focus should be as less as possible to raise the measurement accuracy. Moreover, it is easier to achieve accurate results at lower gas flow.
Keywords:dry powder inhaler  dispersion effect  single frame and single exposure image method  powder