首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Consistency Conditions for Cone-Beam CT Data Acquired with a Straight-Line Source Trajectory
Authors:LEVINEMargoS  SIDKYEmilY  PANXiaochuan
Institution:LEVINE+Margo+S,SIDKY+Emil+Y,PAN+Xiaochuan+Department+of+Radiology,University+of+Chicago,5841+S.Maryl,+Ave.,Chicago,IL+60637,USA
Abstract:A consistency condition is developed for computed tomography(CT) projection data acquired from a straight-line X-ray source trajectory.The condition states that integrals of normalized projection data along detector lines parallel to the X-ray path must be equal.The projection data is required to be untruncated only along the detector lines parallel to the X-ray path,a less restrictive requirement compared to Fourier conditions that necessitate completely untruncated data.The condition is implemented numeri...
Keywords:data redundancy  John's equation  consistency conditions
本文献已被 CNKI 万方数据 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号